[PATCH v2] staging: iio: ad5933: fix order of cycle conditions
From: Luis de Bethencourt
Date: Wed Jun 01 2016 - 15:26:11 EST
Correctly handle the settling time cycles value. The else branch is an
impossible condition, > 1022 in the else branch of > 511. Flipping the order.
Based on the Table 13 at the bottom of Page 25 of the Data Sheet:
http://www.analog.com/media/en/technical-documentation/data-sheets/AD5933.pdf
Signed-off-by: Luis de Bethencourt <luisbg@xxxxxxxxxxxxxxx>
Reviewed-by: Lars-Peter Clausen <lars@xxxxxxxxxx>
---
drivers/staging/iio/impedance-analyzer/ad5933.c | 6 +++---
1 file changed, 3 insertions(+), 3 deletions(-)
diff --git a/drivers/staging/iio/impedance-analyzer/ad5933.c b/drivers/staging/iio/impedance-analyzer/ad5933.c
index 9f43976..170ac98 100644
--- a/drivers/staging/iio/impedance-analyzer/ad5933.c
+++ b/drivers/staging/iio/impedance-analyzer/ad5933.c
@@ -444,10 +444,10 @@ static ssize_t ad5933_store(struct device *dev,
st->settling_cycles = val;
/* 2x, 4x handling, see datasheet */
- if (val > 511)
- val = (val >> 1) | (1 << 9);
- else if (val > 1022)
+ if (val > 1022)
val = (val >> 2) | (3 << 9);
+ else if (val > 511)
+ val = (val >> 1) | (1 << 9);
dat = cpu_to_be16(val);
ret = ad5933_i2c_write(st->client,
--
2.5.1