Re: [PATCH 5/8] counter/ti-eqep: add support for latched position

From: David Lechner
Date: Wed Oct 27 2021 - 11:40:20 EST


On 10/27/21 2:44 AM, William Breathitt Gray wrote:
On Sat, Oct 16, 2021 at 08:33:40PM -0500, David Lechner wrote:
This adds support to the TI eQEP counter driver for a latched position.
This is a new extension that gets the counter count that was recorded
when an event was triggered. A new device-level latch_mode attribute is
added to select the trigger. Edge capture unit support will be needed
to make full use of this, but "Unit timeout" mode can already be used
to calculate high speeds.

The unit timer could also have attributes for latched_time and
latched_period that use the same trigger. However this is not a use
case at this time, so they can be added later if needed.

I see that "latched_count" holds the captured counter count; would this
"latched_time" hold the captured unit timer time? If so, does that mean
setting the latch mode to "Unit timeout" always results in a
"latched_time" equal to 0 (assuming that's when the timeout event
triggers)?


Some `latched_*` attributes will only be useful for one `latched_mode`
selection but not the other.

These latched registers are used to measure speed. There are two ways
to do this. A) measuring the change in position over a fixed time and
B) measuring the change in time for a fixed change in position. So for
A) latched_mode would be set to trigger on timeout and we would use
the latched_position for the measurement. For B) we would set the
latched_mode to trigger on reading the count register and use the
latched_time as the measurement.

...

static struct counter_comp ti_eqep_device_ext[] = {
+ COUNTER_COMP_DEVICE_ENUM("latch_mode", ti_eqep_latch_mode_read,
+ ti_eqep_latch_mode_write, ti_eqep_latch_modes),

It seems more appropriate to move this alongside "latched_count" as
Count extension because this is setting the trigger mode to latch the
respective Count's count. Or does this particular extension also affect
the "latched_time" capture for the unit timer?


In hardware, there are at least 3 registers that get latched that I
recall. They are in different subsystems (main count, unit timer,
edge capture). So as you have guessed, that is the reason for having
the trigger selection at the device level.