Re: [PATCH 3/4] iio: adc: Add support for the GE HealthCare PMC ADC

From: Jonathan Cameron
Date: Tue Oct 01 2024 - 15:24:48 EST


On Tue, 1 Oct 2024 09:46:17 +0200
Herve Codina <herve.codina@xxxxxxxxxxx> wrote:

> The GE HealthCare PMC Analog to Digital Converter (ADC) is a 16-Channel
> (voltage and current), 16-Bit ADC with an I2C Interface.
>
> Signed-off-by: Herve Codina <herve.codina@xxxxxxxxxxx>

Just one thing to add to David's review.

I'm going to guess this isn't a general purpose ADC? Can you share any info
on what sort of device it is used in?

No problem if not - I'm just curious as I've not seen GE HealthCare I2C parts
before!

> diff --git a/drivers/iio/adc/gehc-pmc-adc.c b/drivers/iio/adc/gehc-pmc-adc.c
> new file mode 100644
> index 000000000000..c46c2fb84d35
> --- /dev/null
> +++ b/drivers/iio/adc/gehc-pmc-adc.c
> @@ -0,0 +1,233 @@


> +
> +static int pmc_adc_read_raw(struct iio_dev *indio_dev, struct iio_chan_spec const *chan,
> + int *val, int *val2, long mask)
> +{
> + struct pmc_adc *pmc_adc = iio_priv(indio_dev);
> + int ret;
> +
> + switch (mask) {
> + case IIO_CHAN_INFO_RAW:
> + ret = pmc_adc_read_raw_ch(pmc_adc, chan->address, val);
> + if (ret)
> + return ret;
> + return IIO_VAL_INT;
> +
> + case IIO_CHAN_INFO_SCALE:
> + *val = 1; /* Raw values are directly read in mV or mA */

Drop this scale and make the channels processed. That saves userspace even applying
the *1 this indicates. Rare to find a device that outputs in our base units
but might as well take advantage of one that does :)

> + return IIO_VAL_INT;
> + }
> +
> + return -EINVAL;
> +}